Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) , Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: The rod is held by the instrument for inserting and guiding into the screw heads via percutaneous delivery during minimally invasive surgery to fuse the spine. Per the surgical technique (Ll554 Pathfinder NXT Surgical Technique);
Enforcement
- Recall Number
- Z-1980-2012
- Event ID
- 62325
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 18, 2012
- Initiation Date
- June 5, 2012
- Classification Date
- July 10, 2012
- Termination Date
- September 3, 2014
- Address
- 345 E Main St, Warsaw, IN, 46580-2746, United States
Description
Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) , Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: The rod is held by the instrument for inserting and guiding into the screw heads via percutaneous delivery during minimally invasive surgery to fuse the spine. Per the surgical technique (Ll554 Pathfinder NXT Surgical Technique);
Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery.
Part numbers 3573-1 (angle handle) all lots including 64PY, 67YA, 69FF, 71GB, 72JL, 73GY
Worldwide Distribution - USA (nationwide) and the countries of: Austria, Australia, Belgium, France, Gelmany, Italy, Spain, Switzerland and the United Kingdom.
249 ( angle handle and straight) in total