The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form an antigen/antibody complex which is measured turbidimetrically.

Recall: Z-1979-2019 · Reported July 24, 2019

Enforcement

Recall Number
Z-1979-2019
Event ID
83095
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Roche Diagnostics Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 24, 2019
Initiation Date
June 3, 2019
Classification Date
July 16, 2019
Termination Date
August 28, 2020
Address
9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States

Description

The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form an antigen/antibody complex which is measured turbidimetrically.

Reason

Calibration failures with Std.E, Sens.E, or Dup.E error due to atypical reaction kinetics and quality control failures have occurred due to contamination of R1 reagent with R3 reagent.

Code Info

Lot 368756

Distribution

The products were distributed to the following US states: AL, IN, and KY.

Quantity

8 units