FDA Enforcement Class II Terminated

CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients.

Recall: Z-1979-2014 · Reported July 9, 2014

Enforcement

Recall Number
Z-1979-2014
Event ID
68459
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Nihon Kohden America Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 9, 2014
Initiation Date
June 3, 2014
Classification Date
July 2, 2014
Termination Date
April 28, 2015
Address
15353 Barranca Pkwy, Irvine, CA, 92618-2216, United States

Description

CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients.

Reason

Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.

Code Info

Model Number: CNS-6201 Software Version 02-26 only. Serial Numbers: 00485, 00488, 00489, 00490, 00491, 00361, 00429, 00430, 00432, 00436, 00443, 00447, 00478, 00486, 00546, 00564, 00573, 00574, 00575, 00576, 00592, 00626, 00638, 00643, 00651, 00674, 00675, 00677, 00678, 00679, 00687.

Distribution

Worldwide Distribution - US in the state of Oklahoma and in the countries of Bolivia, Chile, Germany, Japan, Korea, Taiwan, Yemen, and South Africa.

Quantity

430