FDA Enforcement Class II Ongoing

Hemospray Endoscopic Hemostat: Reference Part Numbers: a) HEMO-7, REF G56572; b) HEMO-7-EU, REF G24663;

Recall: Z-1978-2023 · Reported June 28, 2023

Enforcement

Recall Number
Z-1978-2023
Event ID
92345
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Wilson-Cook Medical Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 28, 2023
Initiation Date
May 22, 2023
Classification Date
June 21, 2023
Address
4900 Bethania Station Rd, Winston Salem, NC, 27105-1203, United States

Description

Hemospray Endoscopic Hemostat: Reference Part Numbers: a) HEMO-7, REF G56572; b) HEMO-7-EU, REF G24663;

Reason

There are potential risks of the Hemospray powder adhering to the distal end of the endoscope, which can result in adhesion of the endoscope to tissue and consequent difficulty or inability to maneuver/remove the endoscope.

Code Info

UDI/DI : a) 00827002565722, b) 00827002246638; All unexpired lot numbers

Distribution

US Nationwide - Worldwide Distribution

Quantity

63,830 units