FDA Enforcement Class I Ongoing

Halyard HEART CATH, SELF REGIONAL kit. Model Number: SELF131-05.

Recall: Z-1975-2026 · Reported May 13, 2026

Enforcement

Recall Number
Z-1975-2026
Event ID
98686
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
AVID Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
May 13, 2026
Initiation Date
March 13, 2026
Classification Date
May 1, 2026
Address
9000 Westmont Dr, Toano, VA, 23168-9351, United States

Description

Halyard HEART CATH, SELF REGIONAL kit. Model Number: SELF131-05.

Reason

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Code Info

Model Number: SELF131-05. UDI-DI (Kit Lot Numbers): 10809160408650 (1611688, 1611689, 1618351, 1621441, 1627930, 1635908, 1644501, 1646021, 1651957, 1654887, 1656582, 1656581, 1659713, 1670884, 1670885, 1671950).

Distribution

US distribution to FL, IL, MO, MS, NC, NE, TX.

Quantity

2,932 kits