FDA Enforcement Class II Ongoing

Serrated Blade Applicator Kit (12 Sample), REF: 552687, For use with the Spife Nexus system, Contents: 30 Blade Applicators. used in invitro diagnostics

Recall: Z-1975-2023 · Reported June 28, 2023

Enforcement

Recall Number
Z-1975-2023
Event ID
92297
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Helena Laboratories, Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
June 28, 2023
Initiation Date
May 2, 2023
Classification Date
June 16, 2023
Address
1530 Lindbergh Dr, Beaumont, TX, 77707-4131, United States

Description

Serrated Blade Applicator Kit (12 Sample), REF: 552687, For use with the Spife Nexus system, Contents: 30 Blade Applicators. used in invitro diagnostics

Reason

Packaging for a serrated blade applicator kit (12 Sample) may contain a serrated blade applicator (18 sample). This results in a characteristic pattern of missed or light samples on the gel, leading to incorrect or no result, or delay in result.

Code Info

Lot # 3-22, UDI-DI: M5255526870

Distribution

Distribution to US states of PA, NE, MI, FL, AR, NY, TX, and CA, and to Thailand.

Quantity

32 units