FDA Enforcement
Class I
Ongoing
Halyard CATH LAB kit. Model Number: SACL75-01.
Recall: Z-1974-2026
·
Reported May 13, 2026
Enforcement
- Recall Number
- Z-1974-2026
- Event ID
- 98686
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- AVID Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- May 13, 2026
- Initiation Date
- March 13, 2026
- Classification Date
- May 1, 2026
- Address
- 9000 Westmont Dr, Toano, VA, 23168-9351, United States
Description
Halyard CATH LAB kit. Model Number: SACL75-01.
Reason
Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Code Info
Model Number: SACL75-01. UDI (Kit Lot Numbers): (01)20809160481377(17)280520(10)341730(30)004(241)UINR44-01 (341730), (01)20809160481377(17)280623(10)341742(30)004(241)SACL75-01 (341742), (01)20809160481377(17)280609(10)345855(30)004(241)SACL75-01 (345855), (01)20809160481377(17)280609(10)345863(30)004(241)SACL75-01 (345863).
Distribution
US distribution to FL, IL, MO, MS, NC, NE, TX.
Quantity
172 kits