FDA Enforcement Class II Ongoing

BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP, REF: 10012241; BD TEXIUM (China), REF: 10012241 CHINA

Recall: Z-1974-2025 · Reported June 18, 2025

Enforcement

Recall Number
Z-1974-2025
Event ID
96737
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
BD SWITZERLAND SARL
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 18, 2025
Initiation Date
May 12, 2025
Classification Date
June 12, 2025
Address
Terre Bonne Park A4, Route De Crassier 17, Eysins, N/A, Switzerland

Description

BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP, REF: 10012241; BD TEXIUM (China), REF: 10012241 CHINA

Reason

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

Code Info

REF/UDI-DI/Lot(Expiration): 10012241/07613203010450/23015679(January 31, 2026), 23035092(March 6, 2026), 23055076(May 4, 2026), 23075108(July 5, 2026), 23095125(September 8, 2026), 23105028(October 3, 2026), 23125147(December 5, 2026), 24015246(January 11, 2027), 24026422(February 27, 2027), 24036125(March 25, 2027), 24045689(April 24, 2027), 24066894(June 25, 2027), 24075112(July 26, 2027), 24095557(September 24, 2027), 24115004(November 1, 2027); 10012241 CHINA/10885403238383/23095707(September 14, 2026), 24025890(February 7, 2027), 24105002(October 3, 2027)

Distribution

Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.

Quantity

1,006,512