SteriPack Sterile Polyester Spun Swabs (25 units/Pack)- Intended for sample collection in the intended nasal area. Catalog Number: 60566RevB
Recall: Z-1970-2023
·
Reported June 21, 2023
Enforcement
- Recall Number
- Z-1970-2023
- Event ID
- 92334
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- LumiraDx
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- June 21, 2023
- Initiation Date
- April 25, 2023
- Classification Date
- June 15, 2023
- Address
- 221 Crescent St, N/A, Waltham, MA, 02453-3475, United States
Description
SteriPack Sterile Polyester Spun Swabs (25 units/Pack)- Intended for sample collection in the intended nasal area. Catalog Number: 60566RevB
Reason
Fails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention to remove part of the swab
Code Info
GTIN-DI: 00850027193205 Pack Lot Number : 85438; Individual Swab Lot Number: 86445 Exp Date: 28-Oct-2023
Distribution
US Nationwide distribution.
Quantity
1003 kits