FDA Enforcement
Class III
Terminated
ADVIA Centaur Calibrator A. Cat. #04800735, SMN 10285904 (6 pack); Cat. # 04800840, SMN 10285905 (6 pack - REF). For calibrating the following assays using the ADVIA Centaur or ACS:180 systems: FT3, T3, T4, Tup, FT4, THEO2.
Recall: Z-1970-2013
·
Reported August 21, 2013
Enforcement
- Recall Number
- Z-1970-2013
- Event ID
- 65626
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 21, 2013
- Initiation Date
- June 28, 2013
- Classification Date
- August 14, 2013
- Termination Date
- July 25, 2014
- Address
- 333 Coney St, East Walpole, MA, 02032-1516, United States
Description
ADVIA Centaur Calibrator A. Cat. #04800735, SMN 10285904 (6 pack); Cat. # 04800840, SMN 10285905 (6 pack - REF). For calibrating the following assays using the ADVIA Centaur or ACS:180 systems: FT3, T3, T4, Tup, FT4, THEO2.
Reason
Calibrator A lot CA77 is under-recovering to the assigned target for the ADVIA Centaur Total T4 Assay.
Code Info
REF 10285904 (6 pack): 89983A77 16 Jul 2013, 90669A77 16 Jul 2013; REF 10285905 (6 pack - REF); 90705A77 16 Jul 2013
Distribution
Worldwide Distribution-Nationwide including PR and the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NV, NY, OH, PA, TN, TX, VA, VT, and WV and the countries of Slovakia, Germany, Turkey, TW, HK, CH, TH, BR, CN, SG, and ZA.
Quantity
265 Cal A 6 packs (US and OUS)