FDA Enforcement Class II Terminated

BD Alaris PCEA Administration Set, Microbore Tubing with Yellow Identification Stripe ASV, Catalog Number 30893-04 (OUS Only) - Product Usage: Microbore Tubing with Yellow Identification Stripe Anti-siphon Valve Use Aseptic Technique Set cannot be used for gravity infusion. 1. Connect PCA syringe to female luer on administration set. 2. Push on syringe plunger to prime set. 3. Refer to PCA module Directions for Use for complete loading and operating instructions. Notes: The set should be changed according to facility protocol or in accordance with currently recognized guidelines for IV therapy. Antisiphon sets cannot be primed by gravity.

Recall: Z-1969-2020 · Reported May 20, 2020

Enforcement

Recall Number
Z-1969-2020
Event ID
85525
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
BD SWITZERLAND SARL
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 20, 2020
Initiation Date
April 9, 2020
Classification Date
May 13, 2020
Termination Date
November 2, 2020
Address
Terre Bonne Park A4, Route De Crassier 17, Eysins, N/A, N/A, Switzerland

Description

BD Alaris PCEA Administration Set, Microbore Tubing with Yellow Identification Stripe ASV, Catalog Number 30893-04 (OUS Only) - Product Usage: Microbore Tubing with Yellow Identification Stripe Anti-siphon Valve Use Aseptic Technique Set cannot be used for gravity infusion. 1. Connect PCA syringe to female luer on administration set. 2. Push on syringe plunger to prime set. 3. Refer to PCA module Directions for Use for complete loading and operating instructions. Notes: The set should be changed according to facility protocol or in accordance with currently recognized guidelines for IV therapy. Antisiphon sets cannot be primed by gravity.

Reason

The products have the potential to leak between the connection of the male luer and the yellow striped tubing. This leak may be observed while priming or during use. The most likely health consequence associated with a leak in an administration set would be inadequate pain control for the patient due to an under-infusion of medication.

Code Info

Lot Numbers: 19016079 19036183 19045426 19045426 19115483 19115483

Distribution

Worldwide distribution - US Nationwide Distribution and the countries of Canada, Belgium.

Quantity

110,350