System 1 Base, 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Enforcement
- Recall Number
- Z-1965-2012
- Event ID
- 61480
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Terumo Cardiovascular Systems Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 18, 2012
- Initiation Date
- June 22, 2012
- Classification Date
- July 9, 2012
- Termination Date
- November 8, 2013
- Address
- 6200 Jackson Road, Ann Arbor, MI, 48103-9586, United States
Description
System 1 Base, 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Terumo Advanced Perfusion System 1 battery charge is being depleted faster than expected during periods of system shutdown. The System 1 may falsely indicate a longer battery life than actual battery time remaining if not subjected to regular use, such as would occur with a backup system. This may result in unexpected system shutdown. Prolonged battery depletion may cause permanent battery damage.
Catalog number: 801763 and serial numbers: 0011-0335, 0337-0610, 1001-1034, 1037-1049, 1100, 1103-1106, 1109-1445, and 1447-1450.
Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV. Internationally to: Republic, Egypt, Germany, Guatamala, Honduras, Hong Kong, India, Indonesia, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam,
approx. 993 units