FDA Enforcement Class II Terminated

Hamilton Medical Infant Flow Sensor, single use, (1.88m), Part number 155500. The infant flow sensor is an accessory for the Hamilton Medical Ventilators. Product Usage: The infant flow sensor is a device to measure patient air flow and pressures within the Hamilton Medical Ventilators.

Recall: Z-1964-2015 · Reported July 8, 2015

Enforcement

Recall Number
Z-1964-2015
Event ID
71434
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hamilton Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 8, 2015
Initiation Date
June 1, 2015
Classification Date
July 2, 2015
Termination Date
August 12, 2015
Address
4990 Energy Way, Reno, NV, 89502-4123, United States

Description

Hamilton Medical Infant Flow Sensor, single use, (1.88m), Part number 155500. The infant flow sensor is an accessory for the Hamilton Medical Ventilators. Product Usage: The infant flow sensor is a device to measure patient air flow and pressures within the Hamilton Medical Ventilators.

Reason

An issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control (APVcmv/CMV+) is used, the ventilator may adjust the ventilation on the inaccurate flow measurement and decrease the amount of ventilation delivered.

Code Info

PN 155500, all lot codes.

Distribution

US Nationwide Distribution.

Quantity

154 total, all part numbers