FDA Enforcement Class II Terminated

Terumo Advanced Perfusion System 1 Electronic O2 Blender/Analyzer. The electronic gas blender provides control and monitoring of the gas flow rate and oxygen content of the gas input to the oxygenator in the perfusion circuit. It also provides monitoring of the carbon dioxide if 95-5 (%oxygen/%carbon dioxide) is used as the blending gas.

Recall: Z-1962-2012 · Reported July 18, 2012

Enforcement

Recall Number
Z-1962-2012
Event ID
61231
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Terumo Cardiovascular Systems Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 18, 2012
Initiation Date
June 22, 2012
Classification Date
July 9, 2012
Termination Date
February 19, 2013
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586, United States

Description

Terumo Advanced Perfusion System 1 Electronic O2 Blender/Analyzer. The electronic gas blender provides control and monitoring of the gas flow rate and oxygen content of the gas input to the oxygenator in the perfusion circuit. It also provides monitoring of the carbon dioxide if 95-5 (%oxygen/%carbon dioxide) is used as the blending gas.

Reason

Terumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local flow control knob.

Code Info

catalog number: 801188 and Serial numbers: 0001-0006, 0008-00050, 00052-00123,00125-00152, 00154-0174,00176-00211, 00213-00241, 00245-00250, 00300-00356, 00358-00365, 00367-00449, 00451-00489, 00492-00522, 00524-00531, 00533-00649, 00651-00660, 00662-00825,0826-0830, 00831-01157, and 01160-01164.

Distribution

Worldwide Distribution-USA (nationwide) and the countries of ARGENTINA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam.

Quantity

1095 Units