FDA Enforcement
Class II
Terminated
19CM ATTACHMENT, BLACK MAX
Recall: Z-1959-2017
·
Reported May 10, 2017
Enforcement
- Recall Number
- Z-1959-2017
- Event ID
- 76059
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- The Anspach Effort, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 10, 2017
- Initiation Date
- October 6, 2016
- Classification Date
- May 4, 2017
- Termination Date
- July 26, 2019
- Address
- 4500 Riverside Dr, Palm Beach Gardens, FL, 33410-4235, United States
Description
19CM ATTACHMENT, BLACK MAX
Reason
Supplied Directions for Use (DFU) did not contain a recommended service interval.
Code Info
All serial numbers shipped to customers; DFU 18-0062; Part No. B-BLACK.
Distribution
Worldwide Distribution - US Nationwide in the states of HI, DC including PR and the countries of Guam and Japan
Quantity
41