FDA Enforcement Class II Terminated

CIVCO General Purpose Sensor, Non-sterile reusable general purpose electromagnetic sensor, for use with 3D Guidance medSAFE trackers, Ref. 610-1067. The firm name on the label is CIVCO, Kalona, IA. The kit contains sensor component part number 410-240. The general purpose sensor within the kit is intended to provide physicians with tools for electromagnetic tracking of instruments with respect to image data. The sensors are designed to attach to custom reusable tracking bracket for leading transducers and other medical devices

Recall: Z-1955-2012 · Reported July 18, 2012

Enforcement

Recall Number
Z-1955-2012
Event ID
61793
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Civco Medical Instruments Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
July 18, 2012
Initiation Date
May 7, 2012
Classification Date
July 6, 2012
Termination Date
January 7, 2015
Address
102 1st St S, Kalona, IA, 52247-9589, United States

Description

CIVCO General Purpose Sensor, Non-sterile reusable general purpose electromagnetic sensor, for use with 3D Guidance medSAFE trackers, Ref. 610-1067. The firm name on the label is CIVCO, Kalona, IA. The kit contains sensor component part number 410-240. The general purpose sensor within the kit is intended to provide physicians with tools for electromagnetic tracking of instruments with respect to image data. The sensors are designed to attach to custom reusable tracking bracket for leading transducers and other medical devices

Reason

The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.

Code Info

Lot number M713910

Distribution

Worldwide Distribution - USA including WI, CA, MA, IN, WA, and NJ. Foreign distribution was made to Germany, Canada, Denmark, Hong Kong, Ukraine, Japan, Italy, Korea, and France.

Quantity

9 kits