CIVCO General Purpose Sensor, Non-sterile reusable general purpose electromagnetic sensor for use with 3D Guidance driveBAY, driveBAY 2, trakSTAR, or trakSTAR2 trackers, Ref. 610-1066. The firm name on the label is CIVCO, Kalona, IA. The kit contains sensor component part number 410-240. The general purpose sensor within the kit is intended to provide physicians with tools for electromagnetic tracking of instruments with respect to image data. The sensors are designed to attach to custom reusable tracking bracket for leading transducers and other medical devices
Enforcement
- Recall Number
- Z-1954-2012
- Event ID
- 61793
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Civco Medical Instruments Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- July 18, 2012
- Initiation Date
- May 7, 2012
- Classification Date
- July 6, 2012
- Termination Date
- January 7, 2015
- Address
- 102 1st St S, Kalona, IA, 52247-9589, United States
Description
CIVCO General Purpose Sensor, Non-sterile reusable general purpose electromagnetic sensor for use with 3D Guidance driveBAY, driveBAY 2, trakSTAR, or trakSTAR2 trackers, Ref. 610-1066. The firm name on the label is CIVCO, Kalona, IA. The kit contains sensor component part number 410-240. The general purpose sensor within the kit is intended to provide physicians with tools for electromagnetic tracking of instruments with respect to image data. The sensors are designed to attach to custom reusable tracking bracket for leading transducers and other medical devices
The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.
Lot numbers M436540, M468460, M469640, M495160, M495190, M523420, M524580, M555880, M563330, M577940, M590400, M597880, M643870, M673080, M685650, M708280, M717280, M726110, M726130, M736410, M782740, M799860, M817860, and M840610.
Worldwide Distribution - USA including WI, CA, MA, IN, WA, and NJ. Foreign distribution was made to Germany, Canada, Denmark, Hong Kong, Ukraine, Japan, Italy, Korea, and France.
100 kits