FDA Enforcement Class II Terminated

CELLTRACKS ANALYZER II (Software Version 2.5.1 & Below. Product Usage: The CELLTRACKS AUTOPREP System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. Cell analyzers such as the CellTracks Analyzer II, CellSpotter System, flow cytometers or microscopes may be used for cell identification and enumeration. The system is fit for in vitro diagnostic use. In the CellTracks Analyzer II software 2.5.1 release notes Pub. No. J40136EN issued 2012-03-26 page 1, the document states that the edit patient sample information screen now includes a patient information section, a new feature unique to software 2.5.1. This section provides a drop down for patient race and cancer type. There are also separate fields for first and last name for both the physician and patient. There is no mention of the drop down patient ID box on the right side of the user interface which auto-populates previous sample data.

Recall: Z-1950-2013 · Reported August 21, 2013

Enforcement

Recall Number
Z-1950-2013
Event ID
64304
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Veridex, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 21, 2013
Initiation Date
December 28, 2012
Classification Date
August 10, 2013
Termination Date
August 11, 2015
Address
1001 US Route 202, Raritan, NJ, 08869-0606, United States

Description

CELLTRACKS ANALYZER II (Software Version 2.5.1 & Below. Product Usage: The CELLTRACKS AUTOPREP System is a general purpose laboratory instrument used with immunomagnetic reagents that capture and enrich target cells, and labeling reagents that differentiate cells in whole blood. Cell analyzers such as the CellTracks Analyzer II, CellSpotter System, flow cytometers or microscopes may be used for cell identification and enumeration. The system is fit for in vitro diagnostic use. In the CellTracks Analyzer II software 2.5.1 release notes Pub. No. J40136EN issued 2012-03-26 page 1, the document states that the edit patient sample information screen now includes a patient information section, a new feature unique to software 2.5.1. This section provides a drop down for patient race and cancer type. There are also separate fields for first and last name for both the physician and patient. There is no mention of the drop down patient ID box on the right side of the user interface which auto-populates previous sample data.

Reason

Veridex, LLC has received a report that a patient's demographics were incorrectly associated with another patient's identification (PID) number. The purpose of this notification is to inform you of the issue and provide instructions to help prevent an occurrence on your CELLTRACKS AUTOPREP System and CELLTRACKS ANALYZER II.

Code Info

CELLTRACKS ANALYZER II (Software Version 2.5.1 & Below). Product Code: CT1476B (9555).

Distribution

Worldwide Distribution - USA Nationwide and the countries of Australia, Canada, China, India, Japan, Singapore, England, France, Germany, Italy and Spain.

Quantity

US: 70 Units; Ex-US 76 Units