FDA Enforcement Class I Ongoing

Dexcom One Continuous Glucose Monitoring System, SKUs(Part Number): STK-DO-006 (MT27409-1), STK-DO-013 (MT27409-1), STK-DO-103 (MT27409-2), STK-DO-109 (MT27409-2)

Recall: Z-1946-2025 · Reported June 25, 2025

Enforcement

Recall Number
Z-1946-2025
Event ID
96743
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Dexcom, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 25, 2025
Initiation Date
May 12, 2025
Classification Date
June 16, 2025
Address
6340 Sequence Dr, N/A, San Diego, CA, 92121-4356, United States

Description

Dexcom One Continuous Glucose Monitoring System, SKUs(Part Number): STK-DO-006 (MT27409-1), STK-DO-013 (MT27409-1), STK-DO-103 (MT27409-2), STK-DO-109 (MT27409-2)

Reason

Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

Code Info

All Lots/ UDI: STK-DO-006 -00386270003188 STK-DO-013 -00386270002327 STK-DO-103-00386270002969 STK-DO-109 -00386270002792

Distribution

worldwide distribution - US Nationwide and the countries of Andorra, Argentina Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, New Zealand, Norway, Oman, Poland, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom.

Quantity

38,127 units