Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone.
Enforcement
- Recall Number
- Z-1946-2024
- Event ID
- 94748
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Newport Corp
- Voluntary / Mandated
- FDA Mandated
- Initial Notification
- Letter
- Report Date
- June 12, 2024
- Initiation Date
- March 27, 2024
- Classification Date
- June 5, 2024
- Address
- Unknown Street, N/A, Ivine, CA, 92606, United States
Description
Newport s N-LHP-928(j) units if sold as a laser head/power supply set, and N5-LPL-944-080 and N6-LPL-944-080 if sold as a power supply alone.
According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.
P/N N-LHP-928 Set S/N (laser head + power supply): 6905AE, 6907AE, 9023AE, 3148AF, 6899AE 6909AE, 9025AE, 5695AF, 5696AF, 5691AF, 5692AF, 5693AF, 5694AF (9 of them in Irvine warehouse) P/N: N6-LPL-944-080 Power supply S/N: 23060081, 23060079
US Nationwide