FDA Enforcement Class II Ongoing

IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Recall: Z-1945-2024 · Reported June 12, 2024

Enforcement

Recall Number
Z-1945-2024
Event ID
94576
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ion Beam Applications S.A.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 12, 2024
Initiation Date
May 10, 2024
Classification Date
June 3, 2024
Address
Chemin Du Cyclotron 3, N/A, Ottignies-Louvain-La-Neuve, N/A, N/A, Belgium

Description

IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Reason

Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances

Code Info

UDI: SAT140: (01)05404013814015(11)190507(21)SAT140 SAT136: (01)05404013801138(11)201105(21)SAT136 Serial Numbers: SAT136, SAT140 PTS-11 versions from PTS-11.1 and PTS-12 versions, with C230 accelerator

Distribution

Worldwide - US Nationwide distribution in the state of Virginia and the country of India.

Quantity

2 units