FDA Enforcement
Class II
Terminated
iQ200 Series Urine Microscopy Analyzer- All part numbers
Recall: Z-1945-2020
·
Reported May 20, 2020
Enforcement
- Recall Number
- Z-1945-2020
- Event ID
- 85423
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Beckman Coulter Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 20, 2020
- Initiation Date
- March 30, 2020
- Classification Date
- May 8, 2020
- Termination Date
- August 31, 2023
- Address
- 250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States
Description
iQ200 Series Urine Microscopy Analyzer- All part numbers
Reason
A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computer s metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the device s CD-ROM.
Code Info
Catalog Numbers: 00-3322, 700-3325, 700-3326, 700-3345, 700-3347, 700-3370, 700-3375, 700-3300RFB, B48999, C10683, C10684., All part numbers
Distribution
US Distribution: Nationwide Foreign Distribution: Worldwide
Quantity
3576 units