syngo Lab Data Manager, SMN 10800057, 10803189, and 10804573.
Recall: Z-1944-2014
·
Reported July 9, 2014
Enforcement
- Recall Number
- Z-1944-2014
- Event ID
- 68394
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 9, 2014
- Initiation Date
- May 20, 2014
- Classification Date
- June 30, 2014
- Termination Date
- August 10, 2015
- Address
- 500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States
Description
syngo Lab Data Manager, SMN 10800057, 10803189, and 10804573.
Reason
Certain configuration parameters may cause the reference ranges to not hold results as expected and release them to the Laboratory Information System (LIS) that should have been held for manual review.
Code Info
syngo Lab Data Manager Application Software versions VA11B, VA12A, and VA12B
Distribution
Worldwide Distribution. US Nationwide, Canada, Australia, Belgium, Czech Republic, Germany, Denmark, Finland, France, United Kingdom, Italy, Netherlands, Portugal, Slovakia, Singapore, Japan, and Spain.
Quantity
336