FDA Enforcement Class II Terminated

Velocity" Biopsy Valve, 25 units per box, Product Usage: The disposable Velocity biopsy valve is used to cover the opening of the biopsy inlet port of a flexible gastrointestinal endoscope. The Velocity biopsy valve provides access for endoscopic device passage and exchange, helps maintain insufflation, and minimizes leakage of biomaterial from the accessory port throughout the gastrointestinal endoscopic procedure. The Velocity biopsy valve provides a connection for irrigation through the Velocity irrigation pump.

Recall: Z-1943-2015 · Reported July 8, 2015

Enforcement

Recall Number
Z-1943-2015
Event ID
71336
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
US Endoscopy Group Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 8, 2015
Initiation Date
February 17, 2015
Classification Date
June 30, 2015
Termination Date
September 21, 2015
Address
5976 Heisley Rd, Mentor, OH, 44060-1873, United States

Description

Velocity" Biopsy Valve, 25 units per box, Product Usage: The disposable Velocity biopsy valve is used to cover the opening of the biopsy inlet port of a flexible gastrointestinal endoscope. The Velocity biopsy valve provides access for endoscopic device passage and exchange, helps maintain insufflation, and minimizes leakage of biomaterial from the accessory port throughout the gastrointestinal endoscopic procedure. The Velocity biopsy valve provides a connection for irrigation through the Velocity irrigation pump.

Reason

The company has received reports of structural non conformance at the connection between the valve body and the irrigation tube, resulting in separation of the components during use.

Code Info

M/N: 00711140; Lot #'s: 1401267, 1404503, 1408953, 1413362, 1417521, 1420739, 1422288, 1422747, 1424014, and 1500443.

Distribution

Worldwide Distribution - US Nationwide in the states of: CA, CO, IA, MA, OH, TX & WV and the countries of: Brazil, Costa Rica, Croatia, Czech Republic, Egypt, France, Germany, India, Indonesia, Italy, Latvia, Lithuania, Philippines, Poland, Romania, Russian Federation, Slovenia & South Africa.

Quantity

2,173 units