FDA Enforcement Class I Ongoing

VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

Recall: Z-1942-2026 · Reported May 13, 2026

Enforcement

Recall Number
Z-1942-2026
Event ID
98554
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Ventec Life Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Report Date
May 13, 2026
Initiation Date
March 25, 2026
Classification Date
May 1, 2026
Address
22002 26th Ave Se, Bothell, WA, 98021-4903, United States

Description

VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

Reason

Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk

Code Info

UDI-DI (REF/UDI-DI): PRT-01198-000/00855573007914, PRT-01185-000/00855573007877 Package/Device Serial Number (package serial number may include a "P" at the end): 152884, 152885, 152886, 152887, 152888, 152889, 152890, 152891, 152892, 152893, 152895, 152896, 152897, 152898, 152899, 152900, 152901, 152902, 152903, 152904, 152908, 152909, 152910, 152911, 152912, 152913, 152914, 152915, 152916, 152917, 152918, 152919, 152986, 152987, 155236, 156507, 156512, 156515, 156516, 156517, 156524, 156532

Distribution

US: OH, GA, CA, MO

Quantity

42