FDA Enforcement Class II Terminated

Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment wrenches, sterile. The drill bits are 30cm long and made of stainless steel. The drill kits are supplied in a sterile state and are single-use only. Used to drill cranial holes using a stereotactic frame.

Recall: Z-1938-2018 · Reported June 6, 2018

Enforcement

Recall Number
Z-1938-2018
Event ID
80002
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ad-Tech Medical Instrument Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 6, 2018
Initiation Date
December 20, 2017
Classification Date
May 25, 2018
Termination Date
October 12, 2018
Address
400 W Oakview Pkwy, N/A, Oak Creek, WI, 53154-7213, United States

Description

Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment wrenches, sterile. The drill bits are 30cm long and made of stainless steel. The drill kits are supplied in a sterile state and are single-use only. Used to drill cranial holes using a stereotactic frame.

Reason

There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm drill bits from DDK2-2.8-30X Disposable Drill Kits, Lot Number 111745 208140649 and vice versa.

Code Info

Catalog No. DDK2-2.4-30X, Lot No. 0111664 208140649, UDI No. 01008418231011741719080110011174 5208140649; Catalog No. DDK2-2.8-30X, Lot No. 0111745 208140649, UDI No. 01008418231011671719080110011166 4208140649

Distribution

Worldwide Distribution: US (nationwide) to states of: CA, DC, FL, MI, MA, NC, NY, WA, WI; and country of: Canada. Distributed internationally to the United Kingdom.

Quantity

30 devices