FDA Enforcement Class II Ongoing

Protocol Configuration Tool (PCT) software v1.3.0 for Hamilton AutoLys STAR used within FoundationOneCDx (F1CDx) installed on Research Triangle Park laboratory Equipment IDs 01342, 01591, 01989, 01990, 01994, 01995.

Recall: Z-1937-2026 · Reported April 29, 2026

Enforcement

Recall Number
Z-1937-2026
Event ID
98640
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Foundation Medicine, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
April 29, 2026
Initiation Date
March 24, 2026
Classification Date
April 23, 2026
Address
150 2nd St, N/A, Cambridge, MA, 02141-2115, United States

Description

Protocol Configuration Tool (PCT) software v1.3.0 for Hamilton AutoLys STAR used within FoundationOneCDx (F1CDx) installed on Research Triangle Park laboratory Equipment IDs 01342, 01591, 01989, 01990, 01994, 01995.

Reason

Potential for spontaneous closures of Hamilton HxRunControl software during execution of Autolys protocols after deployment of the PCT software for AutoLys Hamilton STAR.

Code Info

Protocol Configuration Tool (PCT) software v1.3.0 for Hamilton AutoLys STAR used within FoundationOneCDx (F1CDx)

Distribution

US distribution to North Carolina and Massachusetts.

Quantity

10 units