FDA Enforcement Class II Terminated

Philips Zenition 70, System code 718133 - Product Usage: The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device.

Recall: Z-1933-2020 · Reported May 20, 2020

Enforcement

Recall Number
Z-1933-2020
Event ID
85473
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips North America, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 20, 2020
Initiation Date
April 10, 2020
Classification Date
May 8, 2020
Termination Date
January 11, 2021
Address
3000 Minuteman Rd, N/A, Andover, MA, 01810-1032, United States

Description

Philips Zenition 70, System code 718133 - Product Usage: The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device.

Reason

When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may be exposed (230 V) and/or the video signal can be lost.

Code Info

Serial Numbers 35 37 56 20 50 2 7 8 10 12 13 47 67 34 19 31 45 86 30 59 51 21 54

Distribution

Worldwide distribution - US Nationwide distribution including in the states of MD and MI. The countries of Austria, Belgium, France, Germany, Netherlands, Norway, South Africa, Spain, Switzerland, and United Kingdom.

Quantity

25 total