FDA Enforcement Class II Ongoing

Intera 1.5T Explorer/Nova Dual Product Number: 781108;

Recall: Z-1932-2025 · Reported June 18, 2025

Enforcement

Recall Number
Z-1932-2025
Event ID
96999
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Philips North America
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 18, 2025
Initiation Date
May 29, 2025
Classification Date
June 9, 2025
Address
222 Jacobs St, Cambridge, MA, 02141-2296, United States

Description

Intera 1.5T Explorer/Nova Dual Product Number: 781108;

Reason

identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.

Code Info

Product Number: 781108; UDI-DI: no UDI; Serial Number: 9025;

Distribution

Worldwide - US Nationwide distribution in the states of AZ, CA, DE, MA, MI, MY, NE, NY, SD and the countries of Australia, Brazil, Ecuador, Germany, Iran, Italy, Japan, Lebanon, Libya, Mexico, Poland, Portugal, Spain.

Quantity

1 unit (OUS only)