Philips Zenition 50, System code 718096 - Product Usage: The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device.
Enforcement
- Recall Number
- Z-1932-2020
- Event ID
- 85473
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips North America, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 20, 2020
- Initiation Date
- April 10, 2020
- Classification Date
- May 8, 2020
- Termination Date
- January 11, 2021
- Address
- 3000 Minuteman Rd, N/A, Andover, MA, 01810-1032, United States
Description
Philips Zenition 50, System code 718096 - Product Usage: The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device.
When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may be exposed (230 V) and/or the video signal can be lost.
Serial Numbers 17 7
Worldwide distribution - US Nationwide distribution including in the states of MD and MI. The countries of Austria, Belgium, France, Germany, Netherlands, Norway, South Africa, Spain, Switzerland, and United Kingdom.
25 total