Achieva XR Product Numbers: (1) 781153, (2) 781253;
Enforcement
- Recall Number
- Z-1928-2025
- Event ID
- 96999
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Philips North America
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 18, 2025
- Initiation Date
- May 29, 2025
- Classification Date
- June 9, 2025
- Address
- 222 Jacobs St, Cambridge, MA, 02141-2296, United States
Description
Achieva XR Product Numbers: (1) 781153, (2) 781253;
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.
Product Numbers: (1) 781153, (2) 781253; UDI-DI: (1) no UDI, (2) no UDI; Serial Numbers: All serial numbers;
Worldwide - US Nationwide distribution in the states of AZ, CA, DE, MA, MI, MY, NE, NY, SD and the countries of Australia, Brazil, Ecuador, Germany, Iran, Italy, Japan, Lebanon, Libya, Mexico, Poland, Portugal, Spain.
6 units (5 US, 1 OUS)