FDA Enforcement Class I Ongoing

CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP), Part Number 0998-00-0800-XX & 0998-UC-0800-XX (excluding 0998-00-0800-83, 0998-UC-0800-83 & 0998-00-0800-75) and cart 0997-00-1179

Recall: Z-1928-2018 · Reported June 6, 2018

Enforcement

Recall Number
Z-1928-2018
Event ID
80003
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Maquet Datascope Corp - Cardiac Assist Division
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Other
Report Date
June 6, 2018
Initiation Date
May 3, 2018
Classification Date
May 31, 2018
Address
1300 Macarthur Blvd, N/A, Mahwah, NJ, 07430-2052, United States

Description

CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP), Part Number 0998-00-0800-XX & 0998-UC-0800-XX (excluding 0998-00-0800-83, 0998-UC-0800-83 & 0998-00-0800-75) and cart 0997-00-1179

Reason

Ingress of fluids into the device can affect various electronic circuit boards which would prevent initiation or continuation of therapy.

Code Info

All Serial Numbers distributed from 03/06/2012 to 04/26/2018

Distribution

The devices were distributed to the following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Guam. The devices were distributed to the following foreign countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Brunei, Bulgaria, Canada, Chile, China, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Ghana, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Netherlands, Nigeria, Norway, Oman, Pakistan, Palestine, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrica, Taiwan, Thailand, Turkey, UAE, UK, Vietnam, Yemen.

Quantity

4300