FDA Enforcement
Class II
Terminated
Siemens Healthcare ADVIA Acid/Base Reagent 1 and 2 Kit containing ADVIA Centaur Acid Reagent bottles, lot number: 454778 Kit Catalog Number: 112219 SMN: 10310026 Intended Use: In vitro diagnostic use. Acid Regent is a component of the ADVIA Centaur and ADVIA Centaur XP instruments
Recall: Z-1927-2014
·
Reported July 2, 2014
Enforcement
- Recall Number
- Z-1927-2014
- Event ID
- 68537
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 2, 2014
- Initiation Date
- June 2, 2014
- Classification Date
- June 26, 2014
- Termination Date
- February 5, 2015
- Address
- 333 Coney St, N/A, East Walpole, MA, 02032-1516, United States
Description
Siemens Healthcare ADVIA Acid/Base Reagent 1 and 2 Kit containing ADVIA Centaur Acid Reagent bottles, lot number: 454778 Kit Catalog Number: 112219 SMN: 10310026 Intended Use: In vitro diagnostic use. Acid Regent is a component of the ADVIA Centaur and ADVIA Centaur XP instruments
Reason
Swollen acid bottles due to a trace metal contamination in the Acid Reagent bottles
Code Info
Kit Lot Number: 454779 Exp. Date: 28 February, 2015
Distribution
Worldwide Distribution - US Nationwide and the countries of Canada, Argentina, China, Colombia, Korea, Malyasia, New Zealand, Mexico, Guatemala, and Singapore.
Quantity
3,839 kits