FDA Enforcement Class I Terminated

Magellan Diagnostics LeadCare Ultra Blood Lead Test Kits Part Number: 80-0010

Recall: Z-1926-2017 · Reported May 24, 2017

Enforcement

Recall Number
Z-1926-2017
Event ID
76991
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Magellan Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 24, 2017
Initiation Date
November 29, 2014
Classification Date
May 17, 2017
Termination Date
February 26, 2022
Address
101 Billerica Ave Bldg 4, North Billerica, MA, 01862-1271, United States

Description

Magellan Diagnostics LeadCare Ultra Blood Lead Test Kits Part Number: 80-0010

Reason

LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.

Code Info

Kit lots: 1309BU,1310AU,1312BU,1404AU,1408BU

Distribution

Nationwide Distribution

Quantity

1016