FDA Enforcement
Class II
Terminated
reSET Mobile App, iOS and Android, provides cognitive behavioral therapy, as part of substance abuse treatment.
Recall: Z-1925-2021
·
Reported June 30, 2021
Enforcement
- Recall Number
- Z-1925-2021
- Event ID
- 88061
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Pear Therapeutics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 30, 2021
- Initiation Date
- June 7, 2021
- Classification Date
- June 21, 2021
- Termination Date
- June 14, 2023
- Address
- 201 Mission St Ste 1450, N/A, San Francisco, CA, 94105-1882, United States
Description
reSET Mobile App, iOS and Android, provides cognitive behavioral therapy, as part of substance abuse treatment.
Reason
Due to a software issue, patients with a urine drug screen (UDS) received access to a contingency management wheel spin for potential rewards regardless of the UDS results. Positive UDS results should not have resulted in access to a wheel spin.
Code Info
UDI Codes: reSET Mobile App, iOS (01) 10851580008088 (10) 1 reSET Mobile App, Android (01) 10851580008095 (10) 1
Distribution
U.S.: AZ, CA, DE, FL, IL, IN, KY, MA, MD, MN, NC, NY, OH, TN, WA, and WI O.U.S.: N/A
Quantity
812 units