Magellan Diagnostics LeadCare Ultra Blood Analyzer Part Number: 80-0010
Recall: Z-1925-2017
·
Reported May 24, 2017
Enforcement
- Recall Number
- Z-1925-2017
- Event ID
- 76991
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Magellan Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 24, 2017
- Initiation Date
- November 29, 2014
- Classification Date
- May 17, 2017
- Termination Date
- February 26, 2022
- Address
- 101 Billerica Ave Bldg 4, North Billerica, MA, 01862-1271, United States
Description
Magellan Diagnostics LeadCare Ultra Blood Analyzer Part Number: 80-0010
Reason
LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.
Code Info
All serial numbers
Distribution
Nationwide Distribution
Quantity
49 units