Brand Name: Powered Laser Surgical Instrument Product Name: Soltive Reusable Laser Fibers Model/Catalog Number: TFL-FBX150R, TFL-FBX200R, TFL-FBX365R, TFL-FBX550R, TFL-FBX940R, TFL-FBX940R-JA (Japan Only), TFL-FBX365R-JA (Japan Only), TFL-FBX200BS-JA (Japan Only), TFL-FBX150R-JA (Japan Only), TFL-FBX200R-JA (Japan Only), TFL-FBX550R-JA (Japan Only); Product Description: SOLTIVE Laser Fibers are delivery devices that transmit laser energy from the laser console to the treatment site through the fiber tip. The laser connector secures the device to the laser.
Enforcement
- Recall Number
- Z-1923-2025
- Event ID
- 96723
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Olympus Corporation of the Americas
- Voluntary / Mandated
- Voluntary: Firm initiated
- Report Date
- June 18, 2025
- Initiation Date
- May 6, 2025
- Classification Date
- June 6, 2025
- Address
- 3500 Corporate Pkwy, Center Valley, PA, 18034-8229, United States
Description
Brand Name: Powered Laser Surgical Instrument Product Name: Soltive Reusable Laser Fibers Model/Catalog Number: TFL-FBX150R, TFL-FBX200R, TFL-FBX365R, TFL-FBX550R, TFL-FBX940R, TFL-FBX940R-JA (Japan Only), TFL-FBX365R-JA (Japan Only), TFL-FBX200BS-JA (Japan Only), TFL-FBX150R-JA (Japan Only), TFL-FBX200R-JA (Japan Only), TFL-FBX550R-JA (Japan Only); Product Description: SOLTIVE Laser Fibers are delivery devices that transmit laser energy from the laser console to the treatment site through the fiber tip. The laser connector secures the device to the laser.
Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.
Lot Code: All lot numbers for the following models: Model UDI: 5 Pack / 1 Piece: TFL-FBX150R UDI: 00821925043893 / 00821925043909 TFL-FBX200R UDI: 00821925043954 / 00821925043961 TFL-FBX365R UDI: 00821925044012 / 00821925044029 TFL-FBX550R UDI: 00821925044050 / 00821925044067 TFL-FBX940R UDI: 00821925044104 / 00821925044098 TFL-FBX940R-JA (Japan Only); TFL-FBX365R-JA (Japan Only); TFL-FBX200BS-JA (Japan Only); TFL-FBX150R-JA (Japan Only); TFL-FBX200R-JA (Japan Only); TFL-FBX550R-JA (Japan Only);
Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.
2,675 units (190 US, 2485 OUS)