FDA Enforcement Class II Terminated

Prismaflex 5.10 US, Product Code 113081 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. " Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more with diseases where removal of plasma components is indicated. " Continuous Renal Replacement Therapy (CRRT) in conjunction with the MARS system to conduct MARS treatments for patients weighing 20 kilograms or more. All treatments administered via the Prismaflex control unit must be prescribed by a physician.

Recall: Z-1923-2018 · Reported May 30, 2018

Enforcement

Recall Number
Z-1923-2018
Event ID
79965
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Baxter Healthcare Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 30, 2018
Initiation Date
April 24, 2018
Classification Date
May 23, 2018
Termination Date
June 22, 2021
Address
1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States

Description

Prismaflex 5.10 US, Product Code 113081 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. " Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more with diseases where removal of plasma components is indicated. " Continuous Renal Replacement Therapy (CRRT) in conjunction with the MARS system to conduct MARS treatments for patients weighing 20 kilograms or more. All treatments administered via the Prismaflex control unit must be prescribed by a physician.

Reason

Baxter Healthcare will be installing new firmware on all Prismaflex control units to address the potential for a small number of these units to exhibit a failure mode with the pump module electronics. The failure mode may result in a Voltage Out of Range malfunction alarm, which causes the device to enter a safe state and become inoperable until it is serviced. Baxter will be releasing new firmware that will prevent the malfunction from occurring.

Code Info

All Lots and Serial Numbers

Distribution

Worldwide distribution: United States (including Puerto Rico), and countries of: Canada, Australia, New Zealand, China, Bangladesh, Japan, Malaysia, Myanmar, Sri Lanka, India, Singapore, Vietnam, Hong Kong, Korea, Taiwan, Indonesia, Bruni, Philippines, Thailand, Argentina, Chile, Costa Rica, Dominican Republic, Haiti, Barbados, Bermuda, Venezuela, Cuba, Mexico, Colombia, Brazil and EMEA.

Quantity

1256 units