FDA Enforcement Class II Terminated

Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo Patient Data Module (PDM). The PDM is a major component of the Merge Hemo system.

Recall: Z-1923-2017 · Reported May 10, 2017

Enforcement

Recall Number
Z-1923-2017
Event ID
76946
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Merge Healthcare, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
May 10, 2017
Initiation Date
March 30, 2015
Classification Date
May 2, 2017
Termination Date
May 5, 2017
Address
900 Walnut Ridge Dr, Hartland, WI, 53029-8347, United States

Description

Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001. The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo Patient Data Module (PDM). The PDM is a major component of the Merge Hemo system.

Reason

The units may display the alarm "sampling line clogged" after the flow zeroing procedure that is conducted automatically 1 hour after start up. This could result in the potential of a small bias added to the CO2 measurement and affect the alarm notification.

Code Info

ISA CO2 Sensor serial numbers: 107349, 107357, 107358, 107363, 107370, 107377, 107390, 107407, 107435, 107476, 107509, 107510, 107556, 107659, 107700, 107707, 107768, 107776, 107831, 107832, 107836, 107839, and 107841.

Distribution

US Distribution was made to medical facilities located in AL, FL, ID, LA, MO, NC, OK, SC, TX, and UT. There was government distribution and no foreign/military distribution.

Quantity

23 sensors (PhaseIn Units)