FDA Enforcement Class II Ongoing

Custom Convenience Kits Pre Op Kit - NS, Part Number AMS14994

Recall: Z-1920-2026 · Reported April 29, 2026

Enforcement

Recall Number
Z-1920-2026
Event ID
98659
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Windstone Medical Packaging, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
April 29, 2026
Initiation Date
March 11, 2026
Classification Date
April 22, 2026
Address
1602 4th Ave N, N/A, Billings, MT, 59101-1521, United States

Description

Custom Convenience Kits Pre Op Kit - NS, Part Number AMS14994

Reason

The product correction was initiated because B Braun is issuing a voluntary Urgent Medical Device Correction for gravity IV administration sets as well as pump administration sets utilized with BBMI s Infusomat Space Large Volume Pump, Outlook Pump and Vista Basic Pump due to potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers or other connections and the ability to prime (occlusion).

Code Info

UDI-DI: B098AMS149940; Lot Numbers: 220039 221423 224226 224793 225009

Distribution

US Nationwide distribution in the states of GA, OR, TN, and WA.

Quantity

380 units