FDA Enforcement Class II Terminated

LifeHealth, IRMA TRUPOINT model 443914, typically used in point-of-care settings. For the in vitro measurement of various critical care analytes in human whole blood. These analytes includes: pO2, pCO2, pH, iCa, Na, K and Cl.

Recall: Z-1919-2015 · Reported July 8, 2015

Enforcement

Recall Number
Z-1919-2015
Event ID
71292
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
LifeHealth, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
July 8, 2015
Initiation Date
April 29, 2015
Classification Date
June 26, 2015
Termination Date
October 20, 2015
Address
2656 Patton Rd, Roseville, MN, 55113-1136, United States

Description

LifeHealth, IRMA TRUPOINT model 443914, typically used in point-of-care settings. For the in vitro measurement of various critical care analytes in human whole blood. These analytes includes: pO2, pCO2, pH, iCa, Na, K and Cl.

Reason

The barometric pressure measurement output of this particular IRMA TRUPoint SN 443914 device will drift over time.

Code Info

44803, 44805, 44807, 44808, 44818, 44830, 44837

Distribution

Worldwide distribution. US in the state of MI; INDONESIA, UNITED ARAB EMIRATES, LITHUANIA, POLAND, and PERU.

Quantity

7