FDA Enforcement Class II Terminated

StoneLight 30 Laser System (NS3000). For use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, and General Surgery.

Recall: Z-1918-2015 · Reported July 8, 2015

Enforcement

Recall Number
Z-1918-2015
Event ID
71346
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
New Star Lasers, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 8, 2015
Initiation Date
May 13, 2015
Classification Date
June 26, 2015
Termination Date
October 28, 2016
Address
9085 Foothills Blvd, Roseville, CA, 95747-7130, United States

Description

StoneLight 30 Laser System (NS3000). For use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, and General Surgery.

Reason

During internal testing a situation was discovered where when changing the settings, a new setting could not be accepted and potentially result in a discrepancy in the energy delivered (higher or lower) from the value displayed on the Graphical User Interface display.

Code Info

Part Number: 0010-9400. Serial numbers: AMVM01, AMVM06, AMVM07, AMVG01, AMVP05, AMVP07, AMVP13, AMVS05, AMVK02, AMVS03, AMVS07, AMVE01, AMVS06, AMVP03, AMVS08. AMVS09, AMVS10, AMVS11, AMVS12, AMVM04, AMVM08, AMVM09, AMVM05, AMVS02, AMVK05, AMVP10, AMVS11, AMVP12, AMVP14, AMVP15, AMVP16, AMVP17, AMVP18. AMVS01, AMVK01, AMVK03, AMVK04, AMVK06, AMVK07, AMVK08, AMVK09, AMVK10, AMVM02, AMVM03, AMVM10, AMVP01, AMVP02, AMVP04, AMVP06, AMVP08, AMVP09, AMVS04.

Distribution

Worldwide Distribution -- US, including the states of CA, IL, AZ, FL, OH, MA, MI, and AL; and the country of the Netherlands.

Quantity

52 units