FDA Enforcement Class II Ongoing

STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010

Recall: Z-1916-2024 · Reported June 5, 2024

Enforcement

Recall Number
Z-1916-2024
Event ID
94546
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Synthes (USA) Products LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 5, 2024
Initiation Date
April 25, 2024
Classification Date
May 28, 2024
Address
1301 Goshen Pkwy, West Chester, PA, 19380-5986, United States

Description

STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010

Reason

Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay

Code Info

UDI-DI: 10886982075192 Lot Numbers: 97P6250, 97P6251, 97P6252

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Quantity

261 units