FDA Enforcement Class II Terminated

DePuy Synthes 6.5 mm Titanium Recon Screws w/T25 Stardrive (70mm, 75mm, 80mm, 85mm, 90mm, 95mm, 105mm); part of the Expert Lateral Femur Nail System and indicated to stabilize Subtrochanteric fractures, Ipsilateral neck/shaft fractures, Femoral shaft fractures, Impending pathologic fractures, and malunions/Non-unions.

Recall: Z-1915-2015 · Reported July 8, 2015

Enforcement

Recall Number
Z-1915-2015
Event ID
71361
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Synthes, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 8, 2015
Initiation Date
May 27, 2015
Classification Date
June 26, 2015
Termination Date
December 18, 2015
Address
1302 Wrights Ln E, West Chester, PA, 19380-3417, United States

Description

DePuy Synthes 6.5 mm Titanium Recon Screws w/T25 Stardrive (70mm, 75mm, 80mm, 85mm, 90mm, 95mm, 105mm); part of the Expert Lateral Femur Nail System and indicated to stabilize Subtrochanteric fractures, Ipsilateral neck/shaft fractures, Femoral shaft fractures, Impending pathologic fractures, and malunions/Non-unions.

Reason

certain lots of the Titanium Recon Screws (6.5mm ) were found to contain the incorrect insert GP0760-A (Synthes Titanium Buttress Locking Plate System), instead of the correct insert GP0706-E (Synthes Titanium Intramedullary Nails and Components).

Code Info

Part Numbers : 04.003.024 04.003.025 04.003.026 04.003.027 04.003.028 04.003.029 04.003.031 with Lot Numbers: 8638958 8638957 8534954 8638955 8556020 8556018 8653032 8653036

Distribution

Nationwide Distribution.

Quantity

276