FDA Enforcement Class II Ongoing

Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V

Recall: Z-1911-2025 · Reported June 11, 2025

Enforcement

Recall Number
Z-1911-2025
Event ID
96727
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Olympus Corporation of the Americas
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 11, 2025
Initiation Date
May 5, 2025
Classification Date
June 5, 2025
Address
3500 Corporate Pkwy, N/A, Center Valley, PA, 18034-8229, United States

Description

Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V

Reason

Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.

Code Info

UDI-DI: 04953170405563. Model Number: TJF-Q190V. All serial numbers.

Distribution

Worldwide Distribution: US (Nationwide) and OUS (International) countries to: Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Benin, Bolivia, Brazil, Canada, Cayman Islands, Chile, Colombia, Costa Rica, Croatia, Denmark, Dominican Republic, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Finland, France, French Guiana, Georgia, Germany, Ghana, Guadeloupe, Guatemala, Honduras, Hong Kong, Iceland, India, Iraq, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Morocco, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Palestinian Authority, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Republic of the Congo, R¿union, Russia, Saudi Arabia, Senegal, Seychelles, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, Uruguay, Vietnam, Yemen.

Quantity

12,195 units