Centricity Universal Viewer 6.0 Is a device that displays medical images and data from various imaging sources, and from other healthcare information sources.
Enforcement
- Recall Number
- Z-1910-2019
- Event ID
- 82998
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- GE Healthcare, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 10, 2019
- Initiation Date
- May 16, 2019
- Classification Date
- July 1, 2019
- Termination Date
- April 14, 2023
- Address
- 3000 N Grandview Blvd, N/A, Waukesha, WI, 53188-1615, United States
Description
Centricity Universal Viewer 6.0 Is a device that displays medical images and data from various imaging sources, and from other healthcare information sources.
There is the possibility of viewing studies directly from the Enterprise Archive or VNA with incorrect patient images because the updated series or study is not present in the archive.
UDI # 840682103800 Model # 2098249-002 Software Version: 6.0.9, 6.0SP9.01, CPACS 6.0 SP9.0.1, UV 6.0 SP9.0.1, UV 6.0 SP9.01.1, UV 6.0 SP9.0.1.2, UV 6.0 SP9.01.3, UV6.0 SP9.01.4, UV 6.0 SP9.01.5, UV 6.0 SP9.0.1.6, UV Version 6.0 SP5.0.4.1, UV Version 6.0 SP9, UV Version 6.0 SP9.0.0.1, UV Version 6.0 SP901, UV Version 6.0 SP9011, and UV Version 6.0 SP902
Worldwide Distribution: US (nationwide) to states of: FL, GA, IL, IN, LA, MA, MO, NJ, NY, OR, PA, SC, SD, TX, and WA; and countries of : Brazil, Canada, Chile, France, French Polynesia, Germany, India, Italy, Japan, Kuwait, Lebanon, Singapore, Spain and United Kingdom.
145 (87 US 103 OUS)