FDA Enforcement Class II Ongoing

Synchro .014"guide wire, 35 cm tip, Catalog: M00313010 Synchro .014 guide wire, 300 cm length, Catalog: M00313310 SYNCHRO .010 WIRE SOFT STRAIGHT 200cm, Catalog: M00316310 SYNCHRO .010 WIRE SOFT STRAIGHT 300cm, Catalog: M00316330

Recall: Z-1909-2024 · Reported June 5, 2024

Enforcement

Recall Number
Z-1909-2024
Event ID
94437
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Stryker Neurovascular
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 5, 2024
Initiation Date
April 23, 2024
Classification Date
May 28, 2024
Address
47900 Bayside Pkwy, Fremont, CA, 94538-6515, United States

Description

Synchro .014"guide wire, 35 cm tip, Catalog: M00313010 Synchro .014 guide wire, 300 cm length, Catalog: M00313310 SYNCHRO .010 WIRE SOFT STRAIGHT 200cm, Catalog: M00316310 SYNCHRO .010 WIRE SOFT STRAIGHT 300cm, Catalog: M00316330

Reason

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.

Code Info

Catalog/UDI-DI/Lot: M00313010/07613252186922/123279, 123364, 123404, 123985, 124214, 124216; M00313310/07613252186946/123494; M00316310/07613252187158/123851; M00316330/07613252187165/124156

Distribution

Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.

Quantity

2859