FDA Enforcement Class II Ongoing

Icelock 125 Ratchet, REF: L-125000; Icelock 600XM Ratchet, REF: L-621200; Icelock 621 Ratchet, REF: L-621000; Icelock 621 Ratchet Adaption Kit, REF: L-621100; Ratchet Lock Body, REF: L-692020

Recall: Z-1906-2023 · Reported June 21, 2023

Enforcement

Recall Number
Z-1906-2023
Event ID
92207
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ossur H / F
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 21, 2023
Initiation Date
May 12, 2023
Classification Date
June 13, 2023
Address
Grjothals 5, N/A, Reykjavik, N/A, N/A, Iceland

Description

Icelock 125 Ratchet, REF: L-125000; Icelock 600XM Ratchet, REF: L-621200; Icelock 621 Ratchet, REF: L-621000; Icelock 621 Ratchet Adaption Kit, REF: L-621100; Ratchet Lock Body, REF: L-692020

Reason

A mechanical lock used to connect and release a locking liner into a prosthetic socket have an attachment pin and lock that can wear down faster than expected, leading to the potential for the pin to disengage from the lock, leading to loss of suspension of the external prosthetic system that replaces a missing lower limb, which can lead to falling and injury.

Code Info

REF/UDI-DI: L-125000/05690967566142; L-621200/05690967596590; L-621000/05690967118334; L-621100/05690967155322; L-692020/05690967128548. Lots: HF210205, HF210519, HF211008, HF220303, MX210518, MX210816, MX210823, MX211022, MX220316.

Distribution

US: SC, FL,NM, PA, VA, CA, TX, MT, UT, OH, NV, KS, MI, GA, IA, WA, TN, AZ, MN, NC, IN, MS, AL, NY, ID, DE, OK, MA, WI, OR, KY, MD, CO, WV, LA, IL, NJ, CT, MO, NE, WY, AR, SD, PR

Quantity

4,190