FDA Enforcement Class II Terminated

Access SARS-CoV-2 lgG II Calibrator, REF C69058, IVD, CE UDI: (01)15099590742751

Recall: Z-1906-2021 · Reported June 30, 2021

Enforcement

Recall Number
Z-1906-2021
Event ID
87960
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
June 30, 2021
Initiation Date
May 7, 2021
Classification Date
June 21, 2021
Termination Date
March 27, 2023
Address
250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States

Description

Access SARS-CoV-2 lgG II Calibrator, REF C69058, IVD, CE UDI: (01)15099590742751

Reason

SARS-CoV-2 lgG II numerical results may potentially be multiplied by a factor of 1000 on systems running with assay protocol file (APF) and access assay file (AAF) versions resulting in falsely elevated numerical values.

Code Info

All Lots

Distribution

US: Alabama, Alaska, Arizona, Arkansas, California, Florida, Georgia, Illinois, Iowa, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Missouri, New Hampshire, New Jersey, New York, North Dakota, Ohio, Oklahoma, Pennsylvania, Puerto Rico, Tennessee, Texas, Virginia, West Virginia, Wisconsin, Wyoming, OUS: Andorra, Austria, Croatia, Czech Republic, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, Israel, Italy, Japan, Jordan, Poland, Portugal, San Marino, Slovakia, Slovenia, Spain, United Arab Emirates

Quantity

3,843 units