FDA Enforcement Class II Completed

MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3

Recall: Z-1905-2025 · Reported June 11, 2025

Enforcement

Recall Number
Z-1905-2025
Event ID
96831
Classification
Class II
Status
Completed
Product Type
Devices
Firm
VIRTUAL INCISION CORPORATION
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 11, 2025
Initiation Date
April 23, 2025
Classification Date
June 4, 2025
Address
1501 Old Cheney Rd, N/A, Lincoln, NE, 68512-2505, United States

Description

MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3

Reason

The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.

Code Info

UDI-DI (01)00850038042028; Lot Numbers: 1020040211, 1020040212, 1020040213

Distribution

US Distribution to States: CA, FL, NE

Quantity